Study profile · Pipek et al. 2021

Comparison of SpO2 and heart rate values on Apple Watch and conventional commercial oximeters devices in patients with lung disease

Strong correlation with commercial pulse oximeters for heart rate (r = 0.995) and oxygen saturation (r = 0.81). Skin colour, wrist circumference, wrist hair and nail polish showed no significant effect.

Pipek LZ, Nascimento RFV, Acencio MMP, Teixeira LR · Scientific Reports 2021 · DOI 10.1038/s41598-021-98453-3 · PubMed 34556765

Fact sheet

Study type
Validation study · peer-reviewed
Sample
100 patients
Population
COPD and interstitial lung disease, outpatient clinic
Reference method
Two commercial finger pulse oximeters
Devices
Apple Watch Series 6
Funding
not stated in the abstract
Extraction
abstract only – full-text details deliberately omitted
Checked on
2026-11-17

Limitations: Reference was finger pulse oximeters, not arterial blood gas; correlation is not an error measure.

Results in the register

DeviceMetricSituationError measure (original)Result
Apple Watch Series 6Oxygen saturationOxygen saturation (SpO2)Restcorrelation r0.81
Apple Watch Series 6Heart rateHeart rateRestcorrelation r0.995

Original source: Pipek LZ, Nascimento RFV, Acencio MMP, Teixeira LR (2021): Comparison of SpO2 and heart rate values on Apple Watch and conventional commercial oximeters devices in patients with lung disease. Scientific Reports.