Study profile · Stone et al. 2021

Assessing the Accuracy of Popular Commercial Technologies That Measure Resting Heart Rate and Heart Rate Variability

According to the authors, all seven products measured mean resting heart rate accurately; the largest deviation was for CameraHRV (MAPE 17.26%). For RMSSD, errors varied widely: from 4.10% (HRV4Training) and 6.84% (OURA) to 112.36% (CameraHRV).

Stone JD, Ulman HK, Tran K, et al. · Frontiers in Sports and Active Living 2021 · DOI 10.3389/fspor.2021.585870 · PubMed 33733234

Fact sheet

Study type
Validation study · peer-reviewed
Sample
5 adults, 148 trials
Population
Healthy young adults, at rest
Reference method
Multi-lead ECG
Devices
Sieben kommerzielle Geräte/Apps, u. a. OURA, HRV4Training, CameraHRV
Funding
not stated in the abstract
Extraction
abstract only – full-text details deliberately omitted
Checked on
2026-11-17

Limitations: Very small sample (5 people); rest only. PPG-based HRV is strictly pulse rate variability (PRV).

Results in the register

DeviceMetricSituationError measure (original)Result
CameraHRV (App)Resting heart rateHeart rateRestMAPE17.26%
HRV4Training (App)Resting heart rateHeart rateRestMAPE2.34%
OURARMSSD at restHRVRestMAPE (CCC)6.84% (CCC 0.91)
HRV4Training (App)RMSSD at restHRVRestMAPE (CCC)4.10% (CCC 0.98)
CameraHRV (App)RMSSD at restHRVRestMAPE (CCC)112.36% (CCC 0.04)