Syrax Wearable Evidence Index

Heart rate: How accurate are wearables? Studies at a glance

Optical heart rate (PPG) is tested against ECG. At rest and during steady exercise it is usually accurate; error increases with changes in intensity. Rest and training values should therefore be considered separately.

5 studies · 6 results

Results

6 results

DeviceMetricSituationError measure (original)ResultStudy
Geräte mit Messung im SchlafHeart rate during sleepHeart rateSleeprelative agreementhighMiller et al. 2022Validation study · abstract + full text
6 von 7 GerätenHeart rateHeart rateExercisemedian error (cycling)< 5 %Shcherbina et al. 2017Validation study · abstract only
Wearables (gepoolt)Heart rate: activity vs restHeart ratemixedabsolute error, relative+30% during activityBent et al. 2020Validation study · abstract only
CameraHRV (App)Resting heart rateHeart rateRestMAPE17.26%Stone et al. 2021Validation study · abstract only
HRV4Training (App)Resting heart rateHeart rateRestMAPE2.34%Stone et al. 2021Validation study · abstract only
Apple Watch Series 6Heart rateHeart rateRestcorrelation r0.995Pipek et al. 2021Validation study · abstract only

Studies in this area

  1. Miller et al. 2022: A Validation of Six Wearable Devices for Estimating Sleep, Heart Rate and Heart Rate Variability in Healthy Adults

    Sensors · Validation study · 53 adults, one night (9 h in bed) · DOI 10.3390/s22166317

    All six devices agreed with polysomnography 86–89% of the time for “asleep or awake”, but only 50–65% for sleep stages. Devices that recorded heart rate during sleep showed high agreement for heart rate and moderate to high agreement for HRV.

  2. Shcherbina et al. 2017: Accuracy in Wrist-Worn, Sensor-Based Measurements of Heart Rate and Energy Expenditure in a Diverse Cohort

    Journal of Personalized Medicine · Validation study · 60 adults (29 men, 31 women) · DOI 10.3390/jpm7020003

    During cycling, six of seven devices achieved a median heart rate error below 5%; errors were largest during walking. No device estimated energy expenditure with an error below 20%. Errors were higher in men, with higher BMI and darker skin.

  3. Bent et al. 2020: Investigating sources of inaccuracy in wearable optical heart rate sensors

    npj Digital Medicine · Validation study · not stated in the abstract · DOI 10.1038/s41746-020-0226-6

    No statistically significant difference in heart rate accuracy across skin tones, but significant differences between devices and activities: absolute error during activity was on average 30% higher than at rest.

  4. Stone et al. 2021: Assessing the Accuracy of Popular Commercial Technologies That Measure Resting Heart Rate and Heart Rate Variability

    Frontiers in Sports and Active Living · Validation study · 5 adults, 148 trials · DOI 10.3389/fspor.2021.585870

    According to the authors, all seven products measured mean resting heart rate accurately; the largest deviation was for CameraHRV (MAPE 17.26%). For RMSSD, errors varied widely: from 4.10% (HRV4Training) and 6.84% (OURA) to 112.36% (CameraHRV).

  5. Pipek et al. 2021: Comparison of SpO2 and heart rate values on Apple Watch and conventional commercial oximeters devices in patients with lung disease

    Scientific Reports · Validation study · 100 patients · DOI 10.1038/s41598-021-98453-3

    Strong correlation with commercial pulse oximeters for heart rate (r = 0.995) and oxygen saturation (r = 0.81). Skin colour, wrist circumference, wrist hair and nail polish showed no significant effect.