Methodology of the Syrax Wearable Evidence Index

How the study register is built: which studies are included, how results are extracted, why there is no device ranking and where the limits are.

Updated: · 13 studies

Question

How accurately do consumer wearables measure sleep, heart rate, HRV, oxygen saturation, steps, energy expenditure and heart rhythm – under which conditions and against which reference method?

Search

Manual search of Europe PMC and PubMed for validation studies, meta-analyses and systematic reviews on consumer wearables. Initial scope: the studies cited in Syrax magazine articles plus current meta-analyses and reviews on sleep measurement from 2025 and 2026.

Metadata (title, authors, journal, year, PMID, PMCID, DOI, funding) come from the Europe PMC API and were checked against the abstracts.

Inclusion

Peer-reviewed studies testing a consumer wearable against a recognised reference: polysomnography (sleep), ECG (heart rate, HRV, rhythm), indirect calorimetry (energy expenditure), hand counting or video (steps), pulse oximetry or blood gas analysis (SpO2).

Meta-analyses and reviews are included, but their pooled values are labelled as such and never attributed to a single device.

Extraction

One row per result: study × device × metric × situation, with the error measure in the study's original metric (agreement, kappa, MAPE, mean difference, sensitivity, specificity, correlation, positive predictive value).

Values are only taken over if they appear in the text read. For “abstract only”, full-text details are deliberately missing – the field stays empty rather than estimated.

Funding and conflicts of interest are stated where given in the abstract or Europe PMC metadata; otherwise “not stated in the abstract”.

No ranking

Agreement, MAPE, mean difference and correlation measure different things and cannot be converted into each other. Studies differ in sample, reference, situation and device generation. The register therefore produces no ranking and no common average across studies.

A result applies to the tested device generation and firmware. Successor models are not automatically rated the same.

HRV and PRV

Wearables measure the pulse wave optically (PPG). Variability calculated from it is strictly pulse rate variability (PRV), not ECG-HRV. The register states the reference (ECG) and the situation for every HRV row.

LEEK Pulse Pro and conflict of interest

Syrax sells the LEEK Pulse Pro. There is no independent validation study for this device; it is therefore not in the register and receives no accuracy rating. Manufacturer claims are not presented here as study results.

Updates and corrections

The register is extended when new studies are published. Every change is listed in the changelog of the overview. Please report errors to the contact address in the imprint.

Licence

The compilation is licensed under CC BY 4.0. Study texts are not copied; only individual figures are cited with a link to the original study (DOI, PubMed).