Study profile · Agostinho et al. 2026
Are Wearable Sleep-Tracking Devices Reliable Alternatives to Polysomnography? A Systematic Review and Meta-Analysis
Wearables overestimated total sleep time and sleep efficiency and underestimated wake, with large variability between devices. No device was consistently superior. Reasonable for global sleep metrics, not suitable as a stand-alone diagnostic tool or for sleep stages.
Agostinho M, Borges M, Pereira T, Borges DF, Soares JI · Behavioral Sleep Medicine 2026 · DOI 10.1080/15402002.2026.2673893 · PubMed 42175611
Fact sheet
- Study type
- Meta-analysis · peer-reviewed
- Sample
- 16 studies, 10 in the meta-analysis
- Population
- Healthy adults
- Reference method
- Polysomnography
- Devices
- Oura Ring (3. Generation), Fitbit (ausgewählte Modelle), weitere
- Funding
- not stated in the abstract
- Extraction
- abstract only – full-text details deliberately omitted
- Checked on
- 2026-11-17
Limitations: Methodological heterogeneity and risk of bias according to the authors.
Results in the register
| Device | Metric | Situation | Error measure (original) | Result |
|---|---|---|---|---|
| gepoolt | Total sleep time, sleep efficiency, wakeSleep | Sleep | Direction of bias | TST/SE over-, WASO underestimated |