Study profile · Miller et al. 2022
A Validation of Six Wearable Devices for Estimating Sleep, Heart Rate and Heart Rate Variability in Healthy Adults
All six devices agreed with polysomnography 86–89% of the time for “asleep or awake”, but only 50–65% for sleep stages. Devices that recorded heart rate during sleep showed high agreement for heart rate and moderate to high agreement for HRV.
Miller DJ, Sargent C, Roach GD · Sensors 2022 · DOI 10.3390/s22166317 · PubMed 36016077
Fact sheet
- Study type
- Validation study · peer-reviewed
- Sample
- 53 adults, one night (9 h in bed)
- Population
- Healthy young adults (25.4 ± 5.9 years), sleep lab in Australia
- Reference method
- Polysomnography (sleep), ECG (heart rate, HRV)
- Devices
- Apple Watch S6, Garmin Forerunner 245 Music, Polar Vantage V, Oura Ring Gen 2, WHOOP 3.0, Somfit
- Funding
- Australian Institute of Sport (per Europe PMC)
- Extraction
- abstract and full text
- Checked on
- 2026-11-17
Limitations: One night per person, young healthy sample. HRV was sampled differently per device (whole night, 4-h periods, 3-min awake test).
Results in the register
| Device | Metric | Situation | Error measure (original) | Result |
|---|---|---|---|---|
| Apple Watch S6 | Sleep/wake detectionSleep | Sleep | Agreement (epochs) | 88 % (κ = 0.30) |
| Garmin Forerunner 245 Music | Sleep/wake detectionSleep | Sleep | Agreement (epochs) | 89 % (κ = 0.35) |
| Polar Vantage V | Sleep/wake detectionSleep | Sleep | Agreement (epochs) | 87 % (κ = 0.44) |
| Oura Ring Gen 2 | Sleep/wake detectionSleep | Sleep | Agreement (epochs) | 89 % (κ = 0.51) |
| WHOOP 3.0 | Sleep/wake detectionSleep | Sleep | Agreement (epochs) | 86 % (κ = 0.44) |
| Somfit | Sleep/wake detectionSleep | Sleep | Agreement (epochs) | 87 % (κ = 0.48) |
| Apple Watch S6 | Sleep stages (wake, light, deep, REM)Sleep | Sleep | Agreement (epochs) | 53 % (κ = 0.20) |
| Garmin Forerunner 245 Music | Sleep stages (wake, light, deep, REM)Sleep | Sleep | Agreement (epochs) | 50 % (κ = 0.25) |
| Polar Vantage V | Sleep stages (wake, light, deep, REM)Sleep | Sleep | Agreement (epochs) | 51 % (κ = 0.28) |
| Oura Ring Gen 2 | Sleep stages (wake, light, deep, REM)Sleep | Sleep | Agreement (epochs) | 61 % (κ = 0.43) |
| WHOOP 3.0 | Sleep stages (wake, light, deep, REM)Sleep | Sleep | Agreement (epochs) | 60 % (κ = 0.44) |
| Somfit | Sleep stages (wake, light, deep, REM)Sleep | Sleep | Agreement (epochs) | 65 % (κ = 0.52) |
| Geräte mit Messung im Schlaf | Heart rate during sleepHeart rate | Sleep | relative agreement | high |
| Geräte mit Messung im Schlaf | HRV (RMSSD) during sleepHRV | Sleep | relative agreement | moderate to highLowest agreement for the 3-min awake HRV test (Garmin). |