Study profile · Stone et al. 2021
Assessing the Accuracy of Popular Commercial Technologies That Measure Resting Heart Rate and Heart Rate Variability
According to the authors, all seven products measured mean resting heart rate accurately; the largest deviation was for CameraHRV (MAPE 17.26%). For RMSSD, errors varied widely: from 4.10% (HRV4Training) and 6.84% (OURA) to 112.36% (CameraHRV).
Stone JD, Ulman HK, Tran K, et al. · Frontiers in Sports and Active Living 2021 · DOI 10.3389/fspor.2021.585870 · PubMed 33733234
Fact sheet
- Study type
- Validation study · peer-reviewed
- Sample
- 5 adults, 148 trials
- Population
- Healthy young adults, at rest
- Reference method
- Multi-lead ECG
- Devices
- Sieben kommerzielle Geräte/Apps, u. a. OURA, HRV4Training, CameraHRV
- Funding
- not stated in the abstract
- Extraction
- abstract only – full-text details deliberately omitted
- Checked on
- 2026-11-17
Limitations: Very small sample (5 people); rest only. PPG-based HRV is strictly pulse rate variability (PRV).
Results in the register
| Device | Metric | Situation | Error measure (original) | Result |
|---|---|---|---|---|
| CameraHRV (App) | Resting heart rateHeart rate | Rest | MAPE | 17.26% |
| HRV4Training (App) | Resting heart rateHeart rate | Rest | MAPE | 2.34% |
| OURA | RMSSD at restHRV | Rest | MAPE (CCC) | 6.84% (CCC 0.91) |
| HRV4Training (App) | RMSSD at restHRV | Rest | MAPE (CCC) | 4.10% (CCC 0.98) |
| CameraHRV (App) | RMSSD at restHRV | Rest | MAPE (CCC) | 112.36% (CCC 0.04) |