Study profile · Shcherbina et al. 2017

Accuracy in Wrist-Worn, Sensor-Based Measurements of Heart Rate and Energy Expenditure in a Diverse Cohort

During cycling, six of seven devices achieved a median heart rate error below 5%; errors were largest during walking. No device estimated energy expenditure with an error below 20%. Errors were higher in men, with higher BMI and darker skin.

Shcherbina A, Mattsson CM, Waggott D, et al. · Journal of Personalized Medicine 2017 · DOI 10.3390/jpm7020003 · PubMed 28538708

Fact sheet

Study type
Validation study · peer-reviewed
Sample
60 adults (29 men, 31 women)
Population
Diverse cohort by age, height, weight, skin tone and fitness (38 ± 11 years)
Reference method
Telemetry ECG (heart rate), indirect calorimetry (energy expenditure)
Devices
Apple Watch, Basis Peak, Fitbit Surge, Microsoft Band, Mio Alpha 2, PulseOn, Samsung Gear S2
Funding
FDA; NLM NIH HHS (per Europe PMC)
Extraction
abstract only – full-text details deliberately omitted
Checked on
2026-11-17

Limitations: Lab conditions; devices from 2015/2016, no longer sold.

Results in the register

DeviceMetricSituationError measure (original)Result
6 von 7 GerätenHeart rateHeart rateExercisemedian error (cycling)< 5 %
alle 7 GeräteEnergy expenditureEnergy expendituremixederrorno device < 20%

Original source: Shcherbina A, Mattsson CM, Waggott D, et al. (2017): Accuracy in Wrist-Worn, Sensor-Based Measurements of Heart Rate and Energy Expenditure in a Diverse Cohort. Journal of Personalized Medicine.