Study profile · Pipek et al. 2021
Comparison of SpO2 and heart rate values on Apple Watch and conventional commercial oximeters devices in patients with lung disease
Strong correlation with commercial pulse oximeters for heart rate (r = 0.995) and oxygen saturation (r = 0.81). Skin colour, wrist circumference, wrist hair and nail polish showed no significant effect.
Pipek LZ, Nascimento RFV, Acencio MMP, Teixeira LR · Scientific Reports 2021 · DOI 10.1038/s41598-021-98453-3 · PubMed 34556765
Fact sheet
- Study type
- Validation study · peer-reviewed
- Sample
- 100 patients
- Population
- COPD and interstitial lung disease, outpatient clinic
- Reference method
- Two commercial finger pulse oximeters
- Devices
- Apple Watch Series 6
- Funding
- not stated in the abstract
- Extraction
- abstract only – full-text details deliberately omitted
- Checked on
- 2026-11-17
Limitations: Reference was finger pulse oximeters, not arterial blood gas; correlation is not an error measure.
Results in the register
| Device | Metric | Situation | Error measure (original) | Result |
|---|---|---|---|---|
| Apple Watch Series 6 | Oxygen saturationOxygen saturation (SpO2) | Rest | correlation r | 0.81 |
| Apple Watch Series 6 | Heart rateHeart rate | Rest | correlation r | 0.995 |